Permeability and Triaxial Shear
Item Description Cells / Panel
Master Panels:
S-500 Permeability / Triaxial Master Panel, 110 V 1
S-50010 Permeability / Triaxial Master Panel, 220 V
S-50033 Permeability / Triaxial Master Panel, 300 psi (CE Compliant) 1
Extension Panel:
S-502 Permeability / Triaxial Extension Panel (S-500 Required) 2
Optional:
442001 10 ml burette assembly
Step 2. Choose your panel(s)
Specications S-500 and S-50010
Air Regulators: (3) Precision bleed-type, 0.1 cfm (0.00005 m³/s), 0.2
psi (1.4 kPa) stability.
(1) Precision bleed-type differential, 0.1 cfm (0.00005
m³/s), 0.2 psi (1.4 kPa) stability.
Burettes: (3) Fitted with 25 ml pipettes, 0.1 ml graduations.
Mechanical Gauges: (1) Vacuum gauge, 0-30 in Hg (0-760 mm Hg),
(1) Supply pressure gauge, 0-300 psi (0-2068 kPa)
Digital Readout: (1) With pressure transducer calibrated to read
0-150 psi (0-1034 kPa), sensitivty to 0.1 psi
(0.7 kPa). Sensitivity 0.2%. Self calibrates every
4 secs. Calibration held in non-volatile memory.
Rotary Valve: (1) 4 positions enables each regulator to be read
individually to allow burette pressure setting. 4th
position for connecting to the S-502 Extension Panel.
Special Valve: Allows lling of annulus of burette, pipette or both
Working Pressure: 150 psi (1034 kPa) maximum
Required Utilities: Air, water, vacuum and drain all provided with ¼-in
tubing connectors located on back panel.
Construction: Durable aluminum frame with sturdy plastic sides.
Hard coated aluminum front panel with screen printed
text. All valves are no-volume change-type.
Dimensions:
(W x D x H)
25 x 7¾ x 37½ in (63.5 x 20 x 95 cm)
Footprint: 25¼ x 15.5 in (64 x 39 cm)
Ship. Vol. 15 ft³ (0.425 m³)
Ship Weight: 70 lb (32 kg)
Specications S-502
Air Regulators: (4) Precision bleed-type, 0.1 cfm (0.00005 m³/s), 0.2
psi (1.4 kPa) stability.
(2) Precision bleed-type differential, 0.1 cfm (0.00005
m³/s), 0.2 psi (1.4 kPa) stability.
Burettes: (6) Fitted with 25 ml pipettes, 0.1 ml graduations
Momentary Valves: (6) Situated above each air regulator to permit
pressure to be set and monitored by the S-500 Digital
Pressure readout.
Manifold Block: 6-way, with tubing to connect to the S-500 Rotary Valve
Special Valve: Allows lling of annulus of burette, pipette or both
Working Pressure: 150 psi (1034 kPa) maximum
Required Utilities: Air, water, vacuum and drain all provided with 1/4-in
tubing connectors located on back panel.
Construction: Durable aluminum frame with sturdy plastic sides.
Hard coated aluminum front panel with screen printed
text. All valves are no-volume change-type.
Dimensions:
(W x D x H)
36.4 x 15.8 x 36.8 in (92.5 x 40 x 93.5 cm)
Ship. Vol. 17 ft³ (0.481 m³)
Ship Weight: 100 lb (45 kg)
S-500 Permeability / Triaxial Master Panel
Note: Select the type and quantity of each panel based on the number of cells
needed for simultaneous testing. One S-500 panel can support up to two S-502
extension panels. The S-502 extension panels cannot operate independently
from the S-500. All DGSI panels are built with quality gauges, valves, and ttings.
Contact DGSI to request a quote to meet your needs.
The DGSI master panel (S-500) and extension panel (S-502) control the pressure
and permeant ow to and from the triaxial or permeability cell. The S-500 contains
the main operation panel which includes air and vacuum gauges to monitor the main
supply, de-airing tank ll, vacuum, and pressure controls, an auxiliary vacuum port,
and a pressure transducer and digital readout with a selector valve for monitoring
pressure to each burette.
Each cell being tested requires three individual burette assembly used to monitor
the cell pressure, sample bottom pedestal, and sample top cap. The S-500 can
supply one cell and the S-502 can supply up to two cells. The S-502 is connected to
the S-500 using tee ttings and tubing from the back of the S-500 panel. Pressure
readings from the S-502 are displayed on the S-500 readout via momentary
switches.
Each burette assembly is constructed with acrylic tube
and top and bottom blocks with stand-off ttings for easy
removal to clean or service as needed. A standard glass
pipette (25 ml with 0.1 ml markings) is supplied inside each
burette assembly. Optional pipettes of 10ml, 5ml, or 1ml
are available upon request. All pipette markings are factory
calibrated.