SOIL-DAQ™
CALIBRATION
Operator’s Manual EZ-DAQ™ version 1.4 22 of 25
LOW READ: Set low displayed reading for low signal (A/D).
HIGH READ:Set high displayed reading for high signal input.
For example, a 1” travel LVDT is actually a little longer, maybe 1.023; Input the actual true
travel the indicator has in the “HIGH READ”. Or for a pressue transducer you would input
the master calibrated reading from the calibration tech’s traceable certified instruments.
Press here and
enter low number.
Usually 0.000.
Press here and enter
high number.